Revision 5
16
August 2021
Figure 1: Model 943-TGS-CE
1.4
Sample Handling System
The Analyzer system utilizes a close coupled mounting arrangement of the measurement cell
block to the tail gas process sampling point to minimize the sample handling system length. The
short length allows for analyzer response times to be optimized while sample transport problems
are mitigated. A small volume of tail gas sample is drawn through the measurement cell by use
of an air-driven
aspirator
where the concentrations of the various species are determined
photometrically.
The measurement sample is drawn from a central point of the process duct using a specifically
sized
sample probe
. After measurement, the process sample (mixed with the aspirator drive
air) is vented to the wall of the process duct. The concentric chamber design of the sample
probe provides both the sampling and venting of the tail gas sample through a single sampling
point on the process duct.
The sample probe incorporates a mechanism for reducing the dew point of sulfur vapour in the
sampled gas stream. Instrument air can be introduced to provide an exposed cool surface to
the incoming sample gas. Excess condensed sulfur is hydrostatically returned to the process
duct. The sulfur vapour concentration is measured and displayed on the analyzer display. The
sulfur vapour concentration is useful for system troubleshooting.
Содержание 943-TGS-CE
Страница 137: ...Revision 5 137 August 2021 Figure 84 Oven Cabinet Details...
Страница 138: ...Revision 5 138 August 2021 Figure 85 Control Cabinet Door Closed...
Страница 139: ...Revision 5 139 August 2021 Figure 86 Control Cabinet Door Removed...
Страница 140: ...Revision 5 140 August 2021 Figure 87 Power Steam Air Signals Connection Details...
Страница 141: ...Revision 5 141 August 2021 Figure 88 AC Wiring Schematic...
Страница 142: ...Revision 5 142 August 2021 Figure 89 DC Signals and Wiring Diagram...
Страница 143: ...Revision 5 143 August 2021 Figure 90 Flow Diagram...
Страница 144: ...Revision 5 144 August 2021 Figure 91 Control and Oven Cabinets Interconnect Wiring...