148
CompuScope 1610C
CompuScope 1610C External Clock
External clocking is a very powerful feature in a digitizer. It allows the user to synchronize the digitizer to an
external system.
It is important to note, however, that the external clock for CompuScope 1610C must be 2 times faster than the
required sample rate, i.e. if 8 MS/s sampling is required, the external clock must be 16 MHz.
A corollary to this is that the “sampling clock” will always be a divide-by-2 of the external clock.
The External Clock must be a TTL signal (0 to +5 Volts) with a maximum frequency of 20 MHz and minimum
frequency of 2 kHz.
The rise and fall times of the clock signal must be better than 8 ns for proper operation at the peak sample rates.
A minimum pulse width of 22.5 ns must be respected.
In other words, the duty cycle of the clock must be between 50% and 55%.
Содержание CompuScope 14100C
Страница 6: ......
Страница 24: ...18 Installing CompactPCI PXI Master Slave multi card systems Notes ...
Страница 30: ...24 Memory organization on CompuScopes except for CS85GC Notes ...
Страница 75: ...Driver installation guide 69 Help About CsTest ...
Страница 76: ...70 Driver installation guide Notes ...
Страница 85: ...CompuScope 85GC 79 CompuScope 85GC simplified block diagram Figure 52 CS85G simplified block diagram ...
Страница 90: ...84 CompuScope 85GC Notes ...
Страница 101: ...CompuScope 82GC 95 CompuScope 82GC simplified block diagram Figure 57 CS82GC simplified block diagram ...
Страница 124: ...118 CompuScope 14100C CompuScope 14100C simplified block diagram Figure 68 CompuScope 14100C simplified block diagram ...
Страница 146: ...140 CompuScope 1610C CompuScope 1610C simplified block diagram Figure 79 CompuScope 1610C simplified block diagram ...
Страница 165: ...CompuScope 3200C 159 CompuScope 3200C simplified block diagram Figure 89 CompuScope 3200C simplified block diagram ...
Страница 174: ...168 Notes Notes ...