E006
Page 24
Figure 20 Time setting menu
Note:
Date and Time settings are logged as Tag event in data records
.
Here, Blinking Segments indicates
current selection.
6.5 Measurement
Starting Measurement
With Default Configuration of the LM-XS Pro E006 data logger, the measurement program will be
started automatically after 10 minutes of device RTC set on power up. The Configuration
parameters are prefixed as per
WHO Specification reference E006/TR06.3 Issue date September
8, 2010,
and user cannot change them from device menu.
➢
The data logger switches to recording mode showing
“REC”
message on display
and “
RUN
”
LED flashing.
➢
Temperature Data is logged at prefixed logging interval of 5 minutes.
Tag Events
➢
When alarm high / low condition occurs in temperature measurement, a special Tag event
with time stamp is logged in data records.