USB-A socket
Various functions can be performed through this USB port such as writing logged data to a
memory stick or updating instrument firmware..
LAN1 (optional)
The LAN1 (Local Area Network) interface allows the instrument to be set-up and controlled
and for results to be collected from a remote computer or PLC. This is unavailable for use
when the RS485 interface is selected. The use of the LAN interface requires detailed network
knowledge and is documented separately from this user guide.
LAN2
The LAN2 interface will be implemented to allow quick transfer of logged data directly to a
PC as well as other types of instrument administration.
R
R
e
e
a
a
r
r
P
P
n
n
e
e
u
u
m
m
a
a
t
t
i
i
c
c
C
C
o
o
n
n
n
n
e
e
c
c
t
t
i
i
o
o
n
n
s
s
-
+
T EST
Pneumatic Outputs
P1 (Jig)
P2 (Vent)
P3
P4
P5
AIR SUPPLY
MIN. 5BAR/MAX. 10BAR
REGULAT OR
SUPPLY
REGULAT ED
AIR IN
REGULAT ED
AIR OUT
"Test -
",
"Test +" Ports
The 'Test -' and Test +' ports are female 1/8" BSP threaded to allow for a variety of connector
types. Connections to these ports must be made using suitable sealing compounds to obtain
leak tight joints. The 'Test +' port is connected to the test item or test jig. It may be necessary
to connect a volume to the 'Test -' port (also known as the reference port) to optimise the test
cycle.
Air Supply Port
External filtering must be provided. See “Safety notes and recommendations”
The air supply port is used to operate the instruments internal valves. It accepts 6mm O/D
tubing. Clean dry air is required at a pressure of between 5bar (72PSI) and 10bar (145PSI).
Regulator Supply
External filtering must be provided. See “Safety notes and recommendations”