22. VPAD R LCP Adaptor table
Maximum guideline values in cP (mPa·s)
RPM
LCP
0.3
21.3K
0.5
12.8K
0.6
10K
1
6.4K
1.5
44.2K
2
3.2K
2.5
2.5K
3
2.1K
4
1.6K
5
1.2K
6
1K
10
640
12
533
20
320
30
213
50
128
60
106
100
64
Table 17.
ATTENTION:
K Indicates thousands.
Example: 7.8K = 7800
M Indicates millions
Example: 1.56M = 1560000
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of the
selected scale.