![Fukuda VaSera VS-2000 Series Скачать руководство пользователя страница 8](http://html.mh-extra.com/html/fukuda/vasera-vs-2000-series/vasera-vs-2000-series_operation-manual_2344837008.webp)
4
Non-Explosion Proof
Using with MRI
Using with Defibrillators
Application to Heart
• Never operate the equipment where flammable gas or fluid such as anesthetic, oxygen and
hydrogen are used.
Explosion or fire may result.
• Never operate the equipment in the presence of flammable anesthetics, high concentration
of oxygen.
This may cause an explosion or fire.
• Never operate the equipment inside a hyperbaric chamber.
This may cause an explosion or fire.
• Do not ground with a gas pipe.
Explosion or fire may result.
• Do not use this product with a magnetic resonance imaging (MRI) device.
This device may be pulled towards the MRI device. The local heating caused by the induced
electromotive force may also cause burns to the patient or cause the performance of this product
to deteriorate.
• Use only the supplied lead cable when defibrillating.
• Do not touch this device during defibrillation.
• This device will return to standard operating mode within 10 seconds. The measurement
accuracy will temporarily decrease during defibrillation, but it will not compromise the
safety of patient and the device.
• Although this device has a type CF applied part, it is not designed to be applied directly to
the heart.
DANGER
WARNING
WARNING
CAUTION
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...