2-3
Chapter 2 Installation
Charging
The battery of this device is charged in the device.
1
Turn OFF the power.
(Refer to "Turning the Power ON and OFF" (P2-12) )
The battery is not charged while the device is operating.
2
Connect the power cable to the device.
(Refer to "Connecting the Power Cable" (P2-8) )
The AC power indicator and charge lamp are lit and
charging starts. The charge lamp indicates the charging
status. (When the optional battery (BTE-001) is installed)
Lit in blue: Charging
Unlit: Charging completed
Lit in yellow: An error occurred during charging
The charge lamp may light in yellow during charging for the following reasons:
• The battery temperature is too high or too low.
• An error occurred in the main unit or battery.
Remove the power cable, wait a while and then charge the battery again.
If the charge lamp still lights in yellow when charging the battery again, contact Fukuda Denshi.
The battery charges from a completely empty state within 2 hours and 30 minutes. The charging time varies depending on
factors such as the level from which the battery is charged and the ambient temperature.
Charging does not start if the battery is nearly full, to prevent overcharging.
It is recommended to charge the battery at or near a normal room temperature (10-30°C).
To ensure that the battery can be used at any time, do not unplug the power cable immediately after turning off the device
when operating the device on AC power. Allow the battery to charge until the charge lamp goes out.
The battery does not charge while the device is being operated.
Below are examples of charging times.
• When the battery temperature is 10°C or above, charging will complete within 2 hours and 30 minutes.
• When the battery temperature is between 0°C and 10°C, charging will complete within 4 hours and 30 minutes.
• When the battery temperature is 0°C or below, or 60°C or above, charging will not be performed.
Charge Lamp
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...