Message
Cause
Countermeasure
Cannot send to host.
The data could not be sent to the DMS (Data
Management System).
When this message is displayed, the data is not sent
to the DMS (Data Management System).
Check the communication settings. If no problem
can be found on the communication settings,
contact your local Fukuda Denshi service
representative.
Cannot write due to
insufficient space on SD
card.
The data could not be saved on the SD card
as the maximum number of recordable data
has been reached.
When this message is displayed, the data is not
recorded. Replace with a new SD card and write the
data again.
Cannot write due to
insufficient space on USB
mem.
The data could not be saved on the USB
memory as the maximum number of
recordable data has been reached.
When this message is displayed, the data is not
saved. Replace with a new USB memory and save
the data again.
Check BP zero
An abnormality has occurred in the pressure
sensor unit
.
• Check whether there is air in the cuff.
• Turn the power OFF and then turn it back ON.
Communication error has
occurred.
An error occurred during communication
with the DMS (Data Management System).
When this message is displayed, the device is not
connected to the DMS (Data Management System).
Contact your local Fukuda Denshi service
representative.
Cuff Error
An error occurred during inflation.
Inspect the cuff and air hose for damage or
breakage.
Message
Cause
Countermeasure
Enter the AF distance.
The AF value is not entered.
• Enter the AF value.
• To enter this value by using the patient's height,
select [Height] for "Distance Input Method"
(PW/BP Examination Settings).
Exc. Infl. Time
The BP measurement timed out and
activated the safety device.
Inspect the cuff and air hose for damages or
breakage.
Exceeded Max P.
Due to abnormal increase of cuff pressure
during the measurement, the safety device
activated.
Inspect the cuff and air hose for damage or
breakage.
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...