13-1
Chapter
13
Appendix
Main Specifications
Display
Color LCD, 1024x768 dots (with LED backlight)
PCG
Frequency characteristic
L filter: 50 Hz (-6 dB/oct)
PWV filter: 165 - 280Hz within -3 dB
Non-Invasive Blood
Pressure Measurement
(BPU-100)
Maximum Measuring range
0 to 300 mmHg
Scale interval
1 mmHg
Pressure accuracy
±3 mmHg
Pressure detection
Semiconductor pressure sensor
Zero balancing
Automatic balancing
Measurement
Oscillometric method
NIBP measurement range
20 to 280 mmHg
Inflation
Automatic inflation by pump
Deflation
Automatic deflation by electromagnetic valve
Safety mechanism
Activates when 330 mmHg or above, or 10 mmHg or above lasts for 130 sec. or
longer
SD Card Slot
Compatible with SD Card Specification 2.0
LAN connector
Complies with IEEE802.3u 100BASE-TX (cable must be 50m or below)
USB ports
Compatible with USB2.0 Full Speed, 3 channels
Safety standard
ANSI/AAMI ES 60601-1:2005
Type of protection against electrical shock
Class I Equipment and Internally Powered Equipment
Degree of protection against electrical shock
NIBP: Type CF applied part, defibrillation-proof applied part
PCG input: Type CF applied part, defibrillation-proof applied part
Power Supply
AC 100 to 240V, 50/60 Hz, 80 VA
DC11.1V (During battery operation)
Dimensions
Main Unit
Approx. 178mm (W) x 181mm (D) x 315mm (H)
4.5kg (weight, excluding battery pack)
Display Unit
Approx. 297mm (W) x 210mm (H) x 95.5mm (thickness) (excluding protrusions)
1.6kg (weight)
BPU-100
Approx. 90mm (W) x 140mm (D) x 43mm (H) 0.40kg (weight)
Operating environment Temperature
10 to 40°C
Humidity
15 to 95% (non-condensing)
Storage environment
Temperature
-10 to +60°C
Humidity
10 to 95% (absolute humidity in conditions below 40°C and 95%RH, excluding
conditions with condensation)
Atmospheric Pressure
800 to 1060hPa
Service life
6 years, provided that periodic maintenance and inspection is made (based on
company standard)
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...