7-2
Chapter 7 Blood Pressure and Pulse Wave Examinations: Other Functions
Correcting the Boundaries for CAVI
Measurement
The boundaries for calculating CAVI can be manually corrected.
The boundaries that can be edited in each examination are as follows.
• CAVI: Second heart sound, right brachial rise, left brachial rise, right brachial notch, left brachial notch, right ankle
rise, left ankle rise
1
Display the waveforms on the
examination result display
2
Double touch the pulse wave to be
changed.
A window for correcting the boundaries is displayed.
3
Change the line marker positions.
1)
Touch the following cursor buttons to move
the line marker positions.
[
→→
]: Moves the cursor a long distance to the right.
[
→
]: Moves the cursor a short distance to the right.
[
←←
]: Moves the cursor a long distance to the left.
[
←
]: Moves the cursor a short distance to the left.
[
□
]: Shows the zoomed in display
Touch this again to return to the normal display. The
zoomed in display can also be shown by touching a
line marker.
Boundaries: If there are two boundaries, one can be
selected by touching it. The selected line marker
changes color.
2)
Touch [
].
4
CAVI, UT, %MAP are recalculated
according to the new positions of the
boundaries and the results are updated.
5
After editing the boundaries, print and
save the data manually.
The CAVI boundaries can also be edited for the examination result loaded from the files. (Refer to "Correcting
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...