5-25
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
6
After selecting the waveform, touch
[Measure].
After the CAVI check display, measurement will start.
7
The waveforms can be also acquired by
pressing the [START] key at step 4, after
the boundaries are displayed, and when
the waveform is stable.
Check the points mentioned in "To Ensure Accurate
Measurement".
The blood pressure measurement starts after the CAVI
check display.
• Measurement will automatically complete after 100
seconds.
• It is also possible to set "Manual" for [PW/BP Exam.]
- [Pulse Exam. Set.] - [Wave Recording Method] in
the Settings.
Simple CAVI Measurement
If measuring only CAVI, the overall examination time can be reduced by setting "ON" for [PW/BP Exam.] -
[Pulse Exam. Set.] - [Simple CAVI Measurement]. Use this function for examinations with 4 channels or
more. Below is the procedure for taking measurements.
The measurement procedure is the same as the standard measurement (P5-6 ). Attach the brachial and
ankle cuffs and PCG microphone. Check the waveform and press the [START] key to start the measurement.
The left brachial blood pressure is measured at the same time as the pulse wave.
• CAVI can be measured in a shorter time if ABI and TBI are not measured.
• If ABI and TBI are also measured, the left brachial blood pressure is measured only once when acquiring the pulse
wave, even if "ON" is set for Inflate Brachial Cuff for TBI.
If this measurement mode is configured so that ABI is not measured, the blood pressure is measured
without recording the left brachial pulse wave during the pulse wave measurement to reduce the time
of CAVI measurement. As a result, the waveform of the left brachial pulse wave will not be recorded
for the color report.
CAVI Measurement
Waveform Acquisition
Press [Start].
Start of
Measurement
End of
Measurement
Examination
Result Display
Save/Send
File
Color Report
(R/L-CAVI)
Left Brachial BP
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...