5-12
Chapter 5 Blood Pressure and Pulse Wave Examinations (CAVI/ABI/TBI): Standard Mode
Waveform Length of Pulse Wave
Select a waveform length for the pulse wave to be obtained in the CAVI examination from 5 seconds, 8
seconds or 16 seconds. A longer waveform length means that measurement takes longer, but it improves
the accuracy as a larger number of pulse waves can be used for the CAVI calculation. Even if pulse waves
that are unreliable for reasons such as arrhythmia are excluded, there will still be sufficient pulse waves that
can be used for CAVI calculation.
Set the waveform length by selecting [PW/BP Exam.] - [Pulse Exam. Set.] - [Pulse Wave Measurement Time]
in the Settings. (Default: 8 sec.)
This function is common to 2-channel, 4-channel and 6-channel examinations.
Real Time CAVI Check
During waveform acquisition, press the [START] key to display a real time pulse wave check (CAVI check).
This function is common to 2-channel, 4-channel and 6-channel examinations. The example shown below is 2-channel
examination.
The CAVI indicator displays the number of usable beats to measure CAVI with a reliable quality. Measure
until it turns green.
Auto Extension
If "Auto" is set for [Pulse Exam. Set.] - [Wave Recording Method] or "ON" is set for [Pulse Exam. Set.] - [Auto
Extension], pulse wave acquisition is automatically extended if the waveform consists of less than 50% of
usable beats due to arrhythmia or other irregular pulse.
Waveform acquisition completes when the waveform consists of 50% or more usable beats.
Auto extension is possible for up to 60 seconds. If a proper waveform cannot be acquired after 60 seconds, auto
extension will cease. In such case, perform the examination again.
CAVI Check Result Display
If "ON" is set for [PW/BP Exam.] - [Pulse Exam. Set.] - [CAVI Check], the CAVI Check display will appear after
the CAVI measurement is completed.
This function is common to 2-channel, 4-channel and 6-channel examinations.
Red
: - -
<24% usable beats
Orange
: -
25-49% usable beats
Yellow
: +
50-74% usable beats
Green
: ++
>75% usable beats
CAVI Indicator
Содержание VaSera VS-2000 Series
Страница 4: ......
Страница 22: ...18 ...
Страница 134: ...6 10 Chapter 6 Blood Pressure and Pulse Wave Examination Simple Mode ...
Страница 140: ...7 6 Chapter 7 Blood Pressure and Pulse Wave Examinations Other Functions ...
Страница 172: ...9 22 Chapter 9 File Transfers Communication History ...
Страница 176: ...10 4 Chapter 10 Recording Daily Reports ...
Страница 214: ...12 12 Chapter 12 Maintenance and Inspection ...
Страница 233: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...