Chapter 14 Troubleshooting
Troubleshooting
Display
Due to noise and myoelectricity
interference, arrhythmia analysis is
suspended.
When "Suspend Arrhy. Analysis during
Noise Interference" is set to [ON] (Initial
Settings > Alarm), arrhythmia analysis
will be suspended at noise and
myoelectricity interference.
The <Cannot analyze> message will be
displayed when the analysis suspended
duration exceeds 30 seconds.
Check the electrode attachment on the bedside
monitor or telemetry transmitter, and remove the noise
source.
Alarm generated on the bedside
monitor can not be silenced or
suspended from this equipment.
[NG] is set for "Alarm Suspend/Alarm
Silence from Central Monitor" (Initial
Settings > Alarm)
Set to [OK].
(
Maintenance Manual "Alarm" P7-2)
HR alarm, PR alarm does not
generate. ON] can not be selected
for the alarm setting.
For the DS-7000 series bedside
monitors, the alarm for the parameter not
selected for the "HR/PR Alarm Source"
(ECG/SpO
2
/BP) on the bedside monitor
will be automatically set to OFF on this
equipment.
Setup cannot be performed on the central monitor.
Alarm indicator does not light.
"Pattern Setup" under "Alarm Indicator"
(Initial Settings > Alarm) is set to [OFF].
Set the alarm indicator flashing pattern for each alarm
level. (Initial Settings > Alarm)
(
Maintenance Manual "Alarm" P7-2)
The [Alarm Silence] key does not
function.
[Disable] is set for "All Beds Alarm
Silence Key". (Initial Settings > Alarm)
When [Disable] is set for "All Beds Alarm Silence Key"
(Initial Settings > Alarm), [Alarm Silence] key for the
fixed key, user key, remote control unit will be
disabled. Set to [Enable] when using the [Alarm
Silence] key.
Situation
Cause
Solution
A certain parameter cannot be
displayed.
The parameter is not set to be displayed
on the bedside monitor.
Set the parameter to be displayed on the bedside
monitor.
The parameter is set to [OFF] on the
"Parameter ON/OFF" screen.
On the "Parameter ON/OFF" screen, select [ON] for
the parameter to be displayed.
(
Waveform and numeric data for
certain bed cannot be displayed.
If using a wired network system, there is
no central monitor with the Central ID,
"001".
Or, the Central ID is duplicated
The central monitor with the Central ID, "001" will
function as a network-administrating monitor and
controls the whole network segment. One of the
central monitors must have the Central ID, "001" in a
network system. Also, make sure not to duplicate the
Central ID with other monitors.
(
Maintenance Manual "Room ID/Central ID
Setup" P2-12)
Monitoring is suspended for that bed.
Press the [Resume] key and resume monitoring for
that patient.
Monitoring is suspended on the bedside
monitor.
Resume monitoring on the bedside monitor.
The bed to be displayed is not selected
under [Bed Sel.]. (Common Setup >
Display Config.)
Select the bed under [Bed Sel.]. (Common Setup >
Display Config.)
(
"Selecting the Displaying Bed" P13-28)
The waveform/numeric data to be
displayed are not selected under
[Waveform] / [Numeric Data]. (Common
Setup > Display Config. > Each Bed)
Select the waveform/numeric data to be displayed.
(
"Numeric Data/Waveform" P13-30)
Situation
Cause
Solution
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......