14-14
Chapter 14 Troubleshooting
Troubleshooting
Bed Register
Alarm
Situation
Cause
Solution
A bed cannot be selected on the
"Bed Register" screen.
The remaining displayable bed is 0.
The maximum numbers of beds that can be
registered are 32 beds.
(
Maintenance Manual "Bed Register" P7-28)
The bed is wired network bed, and
the bed ID or channel ID is not
displayed.
The wired network setup is incorrect.
Check the wired network connection.
The central ID, room/bed ID is incorrect.
Check if the Central ID, Room/Bed ID of this
equipment and the monitors connected to the DS-
LANIII are not duplicated. If duplicated, set the
correct ID.
The Room ID and Bed ID of the
TCON bed are not displayed.
The TCON setting is incorrect.
Check the settings of this equipment and TCON.
A communication error between this
equipment and the bedside monitor can
be considered, or if this equipment is the
remote station, a communication error
with the base station can be considered.
Improve the communication status. The
communication status with the base station can be
verified via the TCON mark
Situation
Cause
Solution
Alarm does not generate.
Alarm is suspended.
Cancel the [Alarm Suspend] under Alarm setup of the
Individual Bed Display.
(
Alarm setup for the parameter is set to
[OFF].
On the alarm setup menu for the corresponding
parameter, set the alarm [ON].
(
"Alarm Limit Setup for Each Parameter" P7-4)
The alarm threshold level is not set for
the parameter.
Set the upper/lower alarm threshold level on the alarm
setup menu for the corresponding parameter.
(
"Alarm Limit Setup for Each Parameter" P7-4)
Alarm sound is not generated.
Alarm sound is suspended.
When the alarm sound is suspended on the DS-8500
system, the alarm sound will be also suspended on
this central monitor.
The alarm sound on this central monitor will resume
when the alarm sound on the DS-8500 system
resumes.
Whether or not to link the alarm sound suspend
function can be set on the "Alarm Setup" under the
"Initial Settings" menu.
(
Maintenance Manual "Alarm" P7-2)
Alarm is silenced.
Press the [Resume All Al. Sound] key on the alarm
setup window of the Individual Bed Display.
Alarm sound is difficult to
recognize.
The volume of the alarm sound is too
low.
If the alarm volume is set too low, alarm occurrence
may not be recognized. On the "Tone/Volume" menu,
increase the alarm volume.
(
Arrhythmia alarm does not
generate. Arrhythmia alarm is not
displayed.
The arrhythmia alarm is set to [OFF].
On the "Arrhythmia Alarm" screen, select [ON] for the
alarm of the corresponding arrhythmia.
(
"Arrhythmia Alarm Setup" P7-6)
The precision of arrhythmia detection
has decreased.
Perform the arrhythmia learn process.
(
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......