7-16
Chapter 7 Alarm Function
Too-Far Alarm
Too-Far Alarm
If [ON] is selected for "Too Far Alarm" on the "Alarm Setup" ("Initial Settings"), and if telemetry transmitter is
outside the transmission range for preprogrammed duration (5 sec. to 60 sec.), <Chk TLM Receive> message will
be displayed and alarm sound will generate in 5 seconds interval.
At this time, event key
will be displayed regardless
of the setting for "Event Key" on the "Alarm Setup"
("Initial Settings").
1
Press the displayed
key.
The alarm sound will be silenced and [Too Far
Alarm Silence] will be displayed.
2
Press the [Too Far Alarm Silence]/[Cancel] key.
[Too Far Alarm Silence]: Too Far alarm will be silenced.
[Cancel] : Too Far alarm will be suspended.
If the too-far condition still remains after the pre-programmed time (duration to generate the alarm: 5 sec. to
60 sec.), alarm will generate again.
When the "SpO
2
Check Sensor" Alarm Occurs
For the telemetry and wired network bed, <SpO
2
Check Sensor> will be displayed when the SpO
2
sensor is detached
from the finger.The displayed color of the message and alarm sound will be according to the alarm level setting.
(
Maintenance Manual "Alarm" P7-2)
1
To Silence the Alarm
Press the event key, or [Individual Alarm Silence] key on the individual bed display.
The alarm will be silenced for the preprogrammed duration.
2
To Cancel the Alarm
Attach the SpO
2
probe to the finger, and perform the measurement properly.
3
<SpO
2
Check Sensor> Alarm Judgment, Message Display, Alarm Sound Setting
"During "Check SpO
2
Sensor" under the "Initial Settings" menu allows the following setting.
Whether or not to perform alarm judgment for SpO
2
measurement during "SpO
2
Check Sensor" alarm
Whether or not to display <SpO
2
Check Sensor> message
Whether or not to generate the alarm sound for <SpO
2
Check Sensor>
(
Maintenance Manual "Alarm" P7-2)
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......