Chapter 7 Alarm Function
General Description
Chapter 7 Alarm Function
General Description
CAUTION
This equipment is provided with the equipment status alarm judgment function, but not
provided with other alarm judgment and arrhythmia analysis function. These results are
transmitted from the bedside monitor or central telemetry receiver.
On a wired network (DS-LANIII), the alarm generated on the bedside monitor will be
transmitted to this equipment with maximum of 5 seconds delay (at NIBP alarm generation).
On a wireless network, the alarm generated on the bedside monitor will be transmitted to
this equipment with maximum of 15 seconds delay.
The adjustable alarm limit increment is different between the DS-8000 series and the DS-
7000 series monitors. Therefore, the set alarm limit may change to the adjustable value
depending on the monitor type constructing the network system.
The alarm messages will be displayed according to the priority. For the same alarm priority,
the alarm message for the newer alarm will be displayed.
For the SV-900, ventilator alarm factor will not be notified to the central monitor.
Depending on the bedside monitor type and software version, the ventilator alarm factor
may not be transmitted to the central monitor.
For details of the bedside monitor type and software version, refer to your nearest service
representative.
The alarm message for the arrhythmia alarm (except Tachy, Brady, Ext Tachy, Ext Brady)
will continue to be displayed for 30 seconds even after the alarm condition dissolves.
During arrhythmia learning, arrhythmia alarm other than Asystole, VF, Pause, Tachy, Brady,
Ext Tachy, Ext Brady will not generate.
If "Suspend Arrhy. Analysis during Noise Interference" ("Initial Settings" > [Alarm Setup]) is
set to [ON], the "Cannot analyze" alarm will generate when analysis is suspended for 30
seconds and longer.
Even when the <Cannot analyze> alarm is generated, alarms for HR, Asystole, VF, Tachy,
Brady, Ext Tachy, Ext Brady will generate.
NOTE
On the full disclosure waveform display, the arrhythmia occurrence point will be displayed 7
seconds before the actual arrhythmia occurrence time.(Excluding Asystole, Tachy, Brady,
Ext Tachy, Ext Brady)
REFERENCE
The alarm sound can be turned ON on the "Tone/Volume" menu.
(
On the "Alarm Setup" under the "Initial Settings" menu, detailed alarm setup can be
performed.
This setup should be performed by our service representative or system administrator of
your institution.
(
Maintenance Manual "Alarm" P7-2)
Содержание DYASCOPE 8000 Series
Страница 1: ......
Страница 2: ......
Страница 12: ...Contents ...
Страница 20: ...Contents ...
Страница 44: ...1 2 Chapter 1 General Description Contents ...
Страница 49: ...2 1 Chapter 2 Name of Parts and Their Functions Contents Chapter 2 Name of Parts and Their Functions ...
Страница 50: ...2 2 Chapter 2 Name of Parts and Their Functions Contents ...
Страница 54: ...2 4 Chapter 2 Name of Parts and Their Functions ...
Страница 56: ...3 2 Chapter 3 Description of the Display Contents ...
Страница 68: ...3 12 Chapter 3 Description of the Display Display on the Extended Display Unit and External Monitor ...
Страница 70: ...4 2 Chapter 4 Basic Operation Contents ...
Страница 82: ...5 2 Chapter 5 Preparation Contents ...
Страница 88: ...5 6 Chapter 5 Preparation Nurse Call Daily Check ...
Страница 90: ...6 2 Chapter 6 Admit Discharge Contents ...
Страница 110: ...7 2 Chapter 7 Alarm Function Contents ...
Страница 138: ...8 2 Chapter 8 Parameter Setup Contents ...
Страница 164: ...8 26 Chapter 8 Parameter Setup Parameter ON OFF ...
Страница 166: ...9 2 Chapter 9 Data Review Contents ...
Страница 190: ...10 2 Chapter 10 Waveform Review Contents ...
Страница 206: ...11 2 Chapter 11 Calculation Contents ...
Страница 212: ...12 2 Chapter 12 Printing Contents ...
Страница 230: ...13 2 Chapter 13 Menu Items Contents ...
Страница 270: ...13 40 Chapter 13 Menu Items Nurse Team Setup ...
Страница 272: ...14 2 Chapter 14 Troubleshooting Contents ...
Страница 296: ...14 24 Chapter 14 Troubleshooting Troubleshooting ...
Страница 298: ...15 2 Chapter 15 Setup Item Default Value Contents ...
Страница 320: ...15 22 Chapter 15 Setup Item Default Value Menu Central Monitor Display ...
Страница 321: ...16 1 Chapter 16 Accessories Contents Chapter 16 Accessories Accessories 16 1 Optional Accessories 16 1 ...
Страница 322: ...16 2 Chapter 16 Accessories Contents ...
Страница 326: ...16 4 Chapter 16 Accessories Optional Accessories ...
Страница 328: ...17 2 Chapter 17 Specification Contents ...
Страница 342: ...17 14 Chapter 17 Specification Technical Information ...
Страница 347: ......
Страница 348: ......
Страница 349: ......