9-6
Chapter 9 Maintenance and Inspection
Electrical Safety Inspection Procedure
Maximum allowable leakage current
The "Type CF" part is applied to this equipment.
WARNING
If value that exceeds the allowable limit is identified during an electrical safety inspection,
immediately stop using the equipment and make appropriate repairs.
Failure to repair the equipment can lead to a serious accident.
Safety Specification of this Equipment
Classification by type of protection: Class I equipment
Classification of applied part by degree of protection: Type CF applied part
Earth Leakage Current
1. Measurement wiring diagram
2. Allowable value
Leakage Current
(Unit: mA)
Type B
Type BF
Type CF
Normal
Condition
Single
Failure
Normal
Condition
Single
Failure
Normal
Condition
Single
Failure
Earth Leakage Current
5
10
5
10
5
10
Touch Current
0.1
0.5
0.1
0.5
0.1
0.5
Patient Leakage Current
(From patient connection to
earth)
Direct
Current
0.01
0.05
0.01
0.05
0.01
0.05
Alternating
Current
0.1
0.5
0.1
0.5
0.01
0.05
Patient Leakage Current
(Current when external voltage
is applied to SIP/SOP)
Direct
Current
0.01
0.05
0.01
0.05
0.01
0.05
Alternating
Current
0.1
0.5
0.1
0.5
0.01
0.05
Patient Leakage Current
(When external voltage is applied to patient
connection of Type F Applied Part)
-
-
-
5
-
0.05
Patient Auxiliary Current
Direct
Current
0.01
0.05
0.01
0.05
0.01
0.05
Alternating
Current
0.1
0.5
0.1
0.5
0.01
0.05
Normal condition
Single failure
5mA
10mA
bl5868
bl5869
Содержание CardiMax 2
Страница 4: ...2 ...
Страница 10: ...Contents ...
Страница 26: ...Chapter 1 General Description Contents ...
Страница 34: ...Chapter 2 Operational Description Contents ...
Страница 42: ...Chapter 3 Wire Connection Diagram Block Diagram Contents ...
Страница 47: ...4 1 Chapter 4 Spare Parts List Contents Chapter 4 Spare Parts List ...
Страница 48: ...Chapter 4 Spare Parts List Contents ...
Страница 52: ...Chapter 5 Disassembly and Assembly Procedure Contents ...
Страница 67: ...6 1 Chapter 6 Software Update Contents Chapter 6 Software Update Software Updating Procedure 6 1 ...
Страница 68: ...Chapter 6 Software Update Contents ...
Страница 70: ...6 2 Chapter 6 Software Update Software Updating Procedure ...
Страница 72: ...Chapter 7 Troubleshooting Contents ...
Страница 80: ...7 8 Chapter 7 Troubleshooting Hardware Trouble ...
Страница 82: ...Chapter 8 Maintenance Contents ...
Страница 104: ...Chapter 9 Maintenance and Inspection Contents ...
Страница 114: ...9 10 Chapter 9 Maintenance and Inspection Replacement of the Accessories ...
Страница 116: ...Chapter 10 Appendix Contents ...
Страница 122: ...10 6 Chapter 10 Appendix Check List ...
Страница 123: ...The company and product names used in this manual are trademarks or registered trademarks of respective companies ...
Страница 124: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Tel 81 3 5684 1455 Fax 81 3 3814 1222 http www fukuda com 4R0101440 201706 ...