Chapter 6
Troubleshooting
6.8 CF Cards
No. Symptom
Check
Remedy
Check if the
icon appears in
the lower right of the screen.
Reinsert the CF card.
Test the CF card in the
maintenance menu. (
☞
5.3.6)
If the test result is OK, the CF card is
defective and should be replaced.
Check the detection signal (pin 10
and 77 on IC61) input to the CF
card connector.
CF card detected: "L"
CF card not detected: "H"
(
☞
10.1.2.C4)
If the CF card is not detected, it is
defective and should be replaced.
1
The error message
"CF card write error"
appears.
Check the +3.3 V power supply
(pin 22 and 43 on IC61) on the CF
card. (
☞
10.1.2.C4)
If the +3.3 V signal is not detected,
FET (Q58) is defective and should be
replaced.
2
The error message
"Data write to the CF
card now" appears.
Wait until the write operation
completes.
If the error message does not go out,
the data may be corrupted. Replace
the CF card. Consult our head office
for information on repair of damaged
cards.
Check if the
icon appears in
the lower right of the screen.
Reinsert the CF card.
3
The error message
"File read error"
appears.
Patient data cannot be found from
the ID. In the file mode, check if
there is any past data for that ID.
Use CF cards containing past data.
Check if the
icon appears in
the lower right of the screen.
Reinsert the CF card.
Test the CF card in the
maintenance menu. (
☞
5.3.6)
If the test result is OK, the CF card is
defective and should be replaced.
Check the detection signal (pin 10
and 77 on IC61) input to the CF
card connector.
CF card detected: "L"
CF card not detected: "H"
(
☞
10.1.2.C4)
If the CF card is not detected, it is
defective and should be replaced.
4
The error message
"CF card read error"
appears.
Check the +3.3 V power supply
(pin 22 and 43 on IC61) on the CF
card. (
☞
10.1.2.C4)
If the +3.3 V signal is not detected,
FET is defective and should be
replaced.
5
The error message
"CF card format
abnormal end. Check
card" appears.
Check if the CF card is one we
recommend.
Replace the CF card with one we
recommend.
Check if the
icon appears in
the lower right of the screen.
Reinsert the CF card.
Test the CF card in the
maintenance menu. (
☞
5.3.6)
If the test result is OK, the CF card is
defective and should be replaced.
Check the detection signal (pin 10
and 77 on IC61) input to the CF
card connector.
CF card detected: "L"
CF card not detected: "H"
(
☞
10.1.2.C4)
If the CF card is not detected, it is
defective and should be replaced.
6
The error message
"File write error"
appears.
Check the +3.3 V power supply
(pin 22 and 43 on IC61) on the CF
card. (
☞
10.1.2.C4)
If the +3.3 V signal is not detected,
FET is defective and should be
replaced.
6-12
Confidential
Copying Prohibited
Содержание VaSera VS-1500
Страница 12: ...Confidential Copying Prohibited x ...
Страница 18: ...0 Chapter 1 Features 1 6 Confidential Copying Prohibited ...
Страница 32: ...Chapter 4 Names of Parts of Equipment 4 6 Confidential Copying Prohibited ...
Страница 42: ...Chapter 5 Maintenance 5 10 Confidential Copying Prohibited Diagonal Line Stripe ...
Страница 178: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 3 ...
Страница 180: ...Chapter 10 Part Location Diagram and Checkpoints Confidential Copying Prohibited 10 5 ...
Страница 191: ...Chapter 10 Part Location Diagram and Checkpoints 10 16 Confidential Copying Prohibited ...
Страница 207: ...Chapter 11 Spare Parts List 11 16 Confidential Copying Prohibited ...
Страница 213: ...Chapter 12 Information of the electromagnetic compatibility EMC 12 6 Confidential Copying Prohibited ...
Страница 216: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 4R0100090 200906 ...