11. Troubleshooting
74
SpO
2
Phenomenon Cause
Solution
The probe size is
improper.
Use a probe, which fit
properly.
The probe is
peeling off or is
affected by the
outside light due to
the poor condition
Attach the probe properly
following the instruction.
SpO
2
value is
unstable.
Transmitting and
measuring LEDs
sensor are dirty.
Clean both LED sensors
from dirt.
■
In Case of Dropping the LX-7230 into Water
In case of dropping the LX-7230 into water containing disinfectant, pick up the
LX-7230 quickly from it. Rinse it well with running water, and dry it thoroughly
with a soft cloth.
CAUTION
Do not use a drier. The LX-7230 shape may change or be broken.
When the LX-7230 is rinsed with running water, make sure to close the
battery compartment lid.
Содержание LX-7230
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Страница 76: ...7 Measurement 56 Blank Page ...
Страница 86: ...9 Other Setting Items 66 Blank Page ...
Страница 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Страница 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Страница 111: ......
Страница 112: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L0105700 20100805 01 ...