9. Other Setting Items
65
■
Changing the Respiration Detection Signal ON/OFF
The default setting of the respiration detection signal is “ON”.
The respiration waveform can be detected when the setting of the respiration
detection signal is turned “ON”.
WARNING
If the LX-7230 is used with minute ventilation rate-adaptive implantable
pacemaker, the respiration detection signal may cause the pacemaker to
pace at its maximum programmed rate. If such event occurs, change the
setting to “OFF” to prevent an occurrence of erroneous pacing rate.
For details of the setting change, contact your local Fukuda Denshi service
representative.
CAUTION
The respiration waveform cannot be measured if the setting of the
respiration detection signal is turned “OFF”.
Make sure to turn OFF the respiration measurement function on the
receiving monitor to prevent an erroneous detection of the respiration
alarm (on the receiving monitor side).
The LCD screen in normal use does not indicate the setting status of the
respiration detection signal ON/OFF. Make sure to take measures, such
as marking on the LX-7230, to distinguish whether the setting of the
respiration detection signal ON/OFF is changed.
■
Changing the LCD Contrast
The LCD display contrast of the LX-7230 can be changed in 8 steps.
For details of the setting change, contact your local Fukuda Denshi service
representative.
Содержание LX-7230
Страница 22: ...1 General Description 2 Blank Page ...
Страница 28: ...3 Preparation 8 Blank Page ...
Страница 36: ...5 Respiration Monitoring 16 Blank Page ...
Страница 64: ...6 SpO2 Monitor LX 7230KM 44 Blank Page ...
Страница 76: ...7 Measurement 56 Blank Page ...
Страница 86: ...9 Other Setting Items 66 Blank Page ...
Страница 88: ...10 Changing the Transmitter Channel and Group ID 68 Blank Page ...
Страница 96: ...12 Cleaning and Disinfection 76 Blank Page ...
Страница 111: ......
Страница 112: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L0105700 20100805 01 ...