Electrosurgery Safety
W A R N I N G
The monitoring system contains protection against interference generated by
electrosurgical instruments. However, operating conditions, surgery site with
respect to the location of ECG electrodes, or the type of instrument used,
may cause noise on the ECG. The noise is generated at the tip of an
electrical knife and is difficult to completely eliminate because of the
frequency components of the ECG. To reduce electrosurgical interference,
take the following precautions:
Location
Locate the electrosurgical unit as far as possible from this unit and the patient
cable. This will help reduce interference on the ECG through the monitor or
cables.
Power Supply
Connect the electrosurgical unit to a power supply that is different from that of the
monitor. This will help prevent interference through the power cable.
Electrode Placement
The amount of interference is considerably different depending on the electrode
position and surgery site. Place the ECG electrodes as far away as possible from
the surgery site. Do not place electrodes in the path between the surgery site and
the ground plate. If the electrodes are placed in this path, the amount of
interference will be quite large. Position (+) and (–) electrodes as close as
possible to each other.
Ground Plate
When using electrosurgical instruments, make sure the contact between the
patient and the ground plate is secure. If the connection is incomplete, the patient
may suffer a burn at the electrode site.
Precautions about Magnetic Resonance Imaging
W A R N I N G
z
Do not operate this equipment in magnetic resonance imaging (MRI)
environments.
z
When conducting MRI test, remove the electrodes and sensors connected
to the patient (test subject).
The local heating caused by the induced electromotive force may cause burn
injury to the patient (subject). For details, refer to the operation manual for the
MRI testing device.
vi
Содержание HS-8000 Series
Страница 4: ...Blank Page ...
Страница 20: ...xvi Blank Page ...
Страница 21: ...General Description Chapter 1 General Description 1 2 Features 1 3 1 1 ...
Страница 24: ...1 4 Blank Page ...
Страница 25: ...Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 4 2 1 ...
Страница 34: ...3 4 Blank Page ...
Страница 62: ...4 28 Blank Page ...
Страница 66: ...HS 8312M Accembly1 HS 8312M 01A 6 2 ...
Страница 67: ...HS 8312M Assembly2 HS 8312M 02A 6 3 ...
Страница 68: ...HS 8312M Assembly3 HS 8312M 03A 6 4 ...
Страница 69: ...HS 8312M Assembly4 HS 8312M 04A 6 5 ...
Страница 70: ...HS 8312N Assembly1 HS 8312N 01A 6 6 ...
Страница 71: ...HS 8312N Assembly2 HS 8312N 02A 6 7 ...
Страница 72: ...6 8 NIBPS 801 Assembly NIBP 801 01A ...
Страница 96: ...8 18 Blank Page ...
Страница 132: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100640 201108 ...