Chapter 5 Initial Settings
Measurement Related Setup
3
The measurement source of BP/NIBP mean value can be selected from waveform or calculation.
[Calc.]: Calculates the mean BP from the following calculation. Mean BP = (Systolic BP + Diastolic BP x 2) /3
[Wave]: In case of NIBP, maximum amplitude of oscillograph will be the mean value (MAP). In case of BP,
the following measurement will be performed.
NOTE
The NIBP specification stated on "Chapter 14 Specification" is based on the validation
when [Wave] is selected.
4
Set the "Arrhythmia Analysis Filter".
[Disp. Waveform]: The filter selected on "Admit/Discharge" menu or "ECG" setup menu will be set.
[Fixed]: The filter will be fixed to 0.5 Hz to 40 Hz.
NOTE
When [Disp. Waveform] is selected, the filter will be set according to the selection on
[Menu > Parameter > ECG]. If [Diag.] is selected, the filter will be 0.5 Hz to 40 Hz which
is the same with [Fixed].
5
Set the "Synchronized Mark/Tone".When [Auto] is selected for "Synchronized Mark/Tone", the priority of the
synchronizing parameter can be set.
[ECG]: The synchronizing priority will be set in the order of ECG>SpO
2
>BP. The synchronized tone will set
to [ON].
[SpO
2
]: The synchronizing priority will be set in the order of SpO
2
>ECG>BP. The synchronized tone will be
set to [ON].
6
Set the display priority of the parameter to be displayed inside the HR/PR numeric data box.
This priority setting will be applied when [Auto] is selected for "HR/PR", or when [HR/PR] user key is used to
switch the HR/PR source.
Select the priority order from the dropdown list.
For example, if [ECG/SpO
2
/BP] is selected, HR/PR source will be set in the priority of ECG>SpO
2
>BP.
7
Set the "Catheter Manufacturer for CC Input".Press one of the keys, and enter the manufacturer name on the
displayed window. (Max. 8 characters)
Example of BP Measurement
1: Systolic BP
2: HR interval
3: Mean BP (MEAN)
4: Diastolic BP
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......