2-3
Chapter 2 Network System Construction
Wired Network System
To validate the DS-LAN setting, it is necessary to restart the system. Make sure to power
cycle the system when the setting is changed.
Select from [DS-LANII (10Mbps) ]/ [DS-LANIII (100Mbps) ].
3
Set the Room ID/Bed ID.
CAUTION
When connected to the wired network, make sure that there are no other bedside
monitors with the same Bed ID. If the Bed ID is duplicated, monitoring on the central
monitor is not possible.
When connecting to the DS-LAN II network, set the ID in the range from 001 to 048.
When connecting to the DS-LAN III network, set the ID in the range from 001 to 100.
NOTE
When connecting to a wired network, set the Room ID/Bed ID.
The set Room ID/Bed ID will be saved on the AC Unit (DSA-81) even after the main power
is turned OFF.The settings may change if the AC Unit is replaced.
Enter the Room ID using the alphanumeric keys, and press the [Regist] key.
The entered ID will be displayed on the upper left of the screen.
Press the entering area for "Bed ID", and enter the ID using the numeric keys.
To display the keys for Room ID again, press the entering area for "Room ID".
Press the [Regist] key. The entered ID will be displayed on the upper left of the screen.
4
Set the "DS-LAN Pat. ID Transmission Start Position".
On the DS-8007 System, patient ID of up to 20 digits can be set, but only 10 digits can be transmitted on a DS-
LAN II network. This setup will set the starting digit from the 20 digits to be transmitted on the DS-LAN II
network.
On the DS-LAN III network, if [Central] is selected for the printer and printing is started on the bedside monitor,
the central monitor printer can print only up to 10 digits.This setup allows to set the starting digit of the 10 digits
to be transmitted.The 10 digits restriction is only for printing, and all 20 digits can be transmitted on the DS-
LAN III network.
Enter the starting position in the range from 1 to 20.
5
Set the "Synchronize Hemodynamic Data with the Central Monitor".(Enabled when the DS-5700 is connected.)
[ON]: 5 latest hemodynamic data will be synchronized between this monitor and the central monitor. Other
hemodynamic data will be deleted.
When the hemodynamic data is edited on this monitor, the result will be also reflected on the central monitor,
and vice versa.
[OFF]: 5 latest data will be transmitted to the central monitor, but the data will not be synchronized between
Содержание Dynascope 8000 Series
Страница 1: ......
Страница 2: ......
Страница 8: ...Contents ...
Страница 14: ...Contents ...
Страница 16: ...ii Safety About the Safety Precautions DSA 81 AC Unit HR 800 Recorder Unit ...
Страница 40: ...1 2 Chapter 1 Installation of the Unit Contents ...
Страница 52: ...2 2 Chapter 2 Network System Construction Contents ...
Страница 62: ...3 2 Chapter 3 Using the External Media Contents ...
Страница 68: ...3 6 Chapter 3 Using the External Media Formatting the Full Disclosure Waveform Card ...
Страница 70: ...4 2 Chapter 4 Connection to the External Devices Contents ...
Страница 86: ...5 2 Chapter 5 Initial Settings Contents ...
Страница 112: ...6 2 Chapter 6 Setup Item Default Value Contents ...
Страница 128: ...7 2 Chapter 7 Replacement Parts Contents ...
Страница 132: ...8 2 Chapter 8 Cleaning Disinfecting Storing Contents ...
Страница 138: ...8 6 Chapter 8 Cleaning Disinfecting Storing Cleaning the Equipment and Sensors ...
Страница 140: ...9 2 Chapter 9 Maintenance Check Contents ...
Страница 153: ......
Страница 154: ......