Ventilator Connection
The DS-7000 system can be connected to a ventilator via serial connector. By connecting a ventilator,
ventilator measurement data can be unified on the patient monitor. Also, ventilator alarm can be
notified to the central monitor via telemetry or wired network.
This section describes the procedure on how to connect the DS-7000 and a ventilator, and to input the
ventilator measurement and alarm.
Ventilator
Ventilator Cable
Servo-i Ventilator
CJ-502
When connecting to a ventilator, check the corresponding software version of the ventilator.
Ventilator
Corresponding Version
Servo-i
v1.5 / v2.0 / v3.0
C A U T I O N
z
The ventilator operation should be performed by well-trained and authorized
personnel.
z
For connecting the DS-7000 and a ventilator, use only the specified
connection cable.
z
Verify that the DS-7000 and the ventilator are properly connected.
z
When connecting the cable, verify that the main power of the DS-7000 and
the ventilator is OFF.
Servo-i
CJ-502
Ventilator Selection
To display the ventilator alarm, serial communication setup is required.
1
Press the Menu
→
Initial Settings
→
Serial Comm. keys.
The serial communication setup menu will be displayed.
Select Servo-i for ventilator.
8
-
24
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...