Precautions for Installing the Equipment
This section describes the environmental condition to use the DS-7000.
C A U T I O N
The installation of this equipment should be performed by our service
representative or a person who is well acquainted with this equipment.
Precautions about the Operating Environment
y
The following environmental conditions should be observed when operating the DS-7000.
・
Surrounding Temperature : 10 to 40
°
C
15 to 35
°
C (with MGU-701/702/Poet IQ)
10 to 35
°
C (with MGU-801P/802/803)
・
Relative Humidity
: 30 to 85% (non-condensing)
y
The DS-7000 is intended for patient monitoring in ICU, CCU, surgery, and ward. Direct use in MRI
environment or home-care should be avoided.
y
The power source should fulfill the following condition.
・
Use a hospital grade 3-way outlet. If a hospital grade outlet is not available, make sure to
connect the equipotential ground terminal with the accessory ground cable.
・
Verify the power voltage and frequency before connecting to an AC power source.
・
Use a power source that can provide adequate power to the device.
y
Pay attention when installing or storing the device. Do not install or store in the following locations.
・
where chemicals are stored or gases may be present
・
where the equipment will be subject to splashing water or humidity from a nebulizer or
vaporizer
・
where the equipment will be subject to direct sunlight
・
Unstable place with inclination, vibration, or shock.
y
Ensure proper ventilation to cool the device.
・
A minimum space of 5 cm is required between vents on the rear side of the monitor and the
wall. If the monitor is embedded in a wall or surrounded by a wall, a minimum space of 10
cm is required.
W A R N I N G
If the monitor is used in an environment not fulfilling the above conditions, not
only the monitor will not deliver its maximum performance, but also damage to
the equipment may occur and safety can not be ensured.
If using in an environment other than specified above, contact our service
representative.
C A U T I O N
Equipotential Grounding
When connecting multiple devices, electrical potential difference may be
generated between the devices. This may result in electric shock to the
patient connected to these devices. Pay special attention for use in Operating
Room, ICU, CCU, Cardiac Catheter Laboratory, and Cardiovascular X-ray
room. To avoid such electrical potential difference, use the accessory ground
cable to connect each device’s potential equalization terminal to the same
ground terminal. This is called equipotential grounding.
8
-
2
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...