W A R N I N G
Objective and constant arrhythmia detection is possible through the fixed
algorithm incorporated in this monitor.
However, excessive waveform morphology change, motion artifact, or the
inability to determine the waveform pattern may cause an error, or fail to make
adequate detection. Therefore, physicians should make final decisions using
manual recording, alarm recording and recall waveform for evaluation.
C A U T I O N
For proper arrhythmia detection and ECG monitoring, verify proper electrode
placement, lead selection, and ECG waveform size. If necessary, turn ON the
AC filter. Improper electrode placement, lead selection, and ECG waveform
size can cause errors in detection.
To Set the Arrhythmia Alarm
ON/OFF of arrhythmia alarm and reference of arrhythmia analysis can be set.
1
Press the Menu
→
Settings
→
Alarm
→
Basic Alarm
→
Arrhythmia keys.
On page 1/2, the alarm setup menu of Asystole, VF, VT, Slow
VT, Tachy, Brady will be displayed.
On page 2/2, the alarm setup menu of Run, Bigeminy,
Trigeminy, Pause, Couplet, Frequent will be displayed.
2
Set the reference range.
Pressing the reference value key will display the
Ï
Ð
keys.
Use the
Ï
Ð
keys to set the reference value.
After setting the reference value, press the Close
key.
7
Arrh
ythmia A
nalysi
s
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...