To Edit the Cardiac Output Data
By performing the CO measurement continuously, mean CO and mean CI can be calculated by editing
the measurement result.
1
Press the Menu
→
Function
→
Cardiac Output
→
Edit keys.
The cardiac output edit menu will be displayed.
The mean CO and mean CI value obtained from the measurement
results will be displayed.
The data can be omitted from the averaging by turning OFF the LED
of the corresponded data.
2
Delete the measurement result.
Pressing the Delete key will delete all the measurement data of
the thermodilution curve with the key LED turned OFF.
3
Input the data to the table.
Pressing the Average CO Input key will input the displayed
mean CO data to the table.
N O T E
If the height, weight, and BSA are changed on the patient admit menu, the
average CI will be recalculated. As the CI will not be recalculated after the
hemodynamics calculation, store the average CI with the hemodynamics
calculation before changing the height, weight, and BSA.
Average Cardiac Output
Average Cardiac Index
Cardiac Output / Cardiac
Index
Thermodilution Curve
Measurement Time / Heart Rate
V Mark: VPC detected during CO measurement.
*
Mark: CO value exceeding the mean CO value
±
10
%
.
7
Cardia
c Outp
ut Measu
rem
ent
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...