7
-
5
●
Display Priority Setup for the Parameter on the Table
The display priority of the parameters on the table can be set.
1
Press the Setup key to display the table setup menu.
2
Set the priority of the parameters to be displayed on the table.
All measured parameters will be displayed on the table.
The display priority is high from the top. On the table
display, the higher priority parameter will be displayed
from the left.
The parameters which priority is not set will be displayed
on the right side on the table display.
3
Select the parameter for each priority level.
The parameter can be selected by directly pressing the
key or using the up and down arrow keys.
The priority level will automatically shift downwards to
allow continuous parameter selection.
N O T E
For apnea data (apnea duration), maximum apnea duration at 1 minute will be stored.
4
Select the numbers of high priority parameters to be set as fixed display.
3 to 6 high priority parameters can be set as fixed display.
Pressing the Fix Xparam. key will sequentially set 3
→
4
→
5
→
6
→
3 for the numbers of parameters.
7
Table/T
rend
Содержание DynaScope 7000 Series
Страница 38: ...xxxvi Blank Page ...
Страница 126: ...4 78 Blank Page ...
Страница 148: ...5 22 Blank Page ...
Страница 259: ...7 15 The Description of the Display Scale Real Time Scale Scale 7 Table Trend ...
Страница 322: ...7 78 Blank Page ...
Страница 428: ...10 32 Blank Page ...
Страница 466: ...11 38 Blank Page ...
Страница 474: ...12 8 Blank Page ...
Страница 475: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4L010175E 201207 ...