Ventilator Connection
By connecting this equipment to a ventilator, ventilator alarm can be notified to the central monitor
through wireless network or TCON system.
This section describes the procedure to connect this unit and ventilator, and to input the ventilator
alarm.
N O T E
Only one ventilator can be connected for each DS-8001 system.
When connecting to a ventilator, check the corresponded software version of the ventilator.
Ventilator Corresponding
Software
Version
SV900 Not
specified
Servo-i
v1.5 / v2.0 / v3.0
Servo-s
v2.0 / v3.0
C A U T I O N
z
The ventilator operation should be performed by a well-trained or
authorized person.
z
When connecting this equipment and the ventilator, use only the
specified connection cable.
z
Verify that this equipment and the ventilator are properly connected.
z
When connecting the cable, verify that the main power of this equipment
and the ventilator are OFF.
Ventilator Connection
1 Turn OFF the power of this equipment and ventilator.
2 Connect one end of the optional ventilator connection cable to the ventilator and the other
side to the serial connector on the left side of this equipment.
Ventilator Connection
Cable
SV-900 CJ-500
Servo-i/Servo-s CJ-502
7
-
17
Содержание DSL-8001
Страница 4: ...Blank Page ...
Страница 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 44: ...2 6 Blank Page ...
Страница 60: ...3 16 Blank Page ...
Страница 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Страница 62: ...Block Wiring Diagram This section describes the block wiring diagram of the DSL 8001 Block Wiring Diagram 4 2 ...
Страница 64: ...4 4 Blank Page ...
Страница 72: ...6 6 Blank Page ...
Страница 102: ...7 30 Blank Page ...
Страница 148: ...11 6 Blank Page ...
Страница 150: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100780 201301 ...