Connection of the HLX-801 Telemetry Transmitter Module
C A U T I O N
Connection of the HLX-801 Telemetry Transmitter Module will be performed
by our service representative. Users should not perform this procedure as
malfunction of the equipment may occur.
1. Turn OFF the power of DSL-8001.
2. Loosen the screw of the maintenance cover on the rear of the DSL-8001, and remove the
maintenance cover.
The removed maintenance cover is not used to attach the HLX-801. Store it because it is used if
the HLX-801 is removed.
3. Connect the connection cable of the DSL-8001 to the HLX-801 connector.
Connection Cable
At this time, pay attention to the connector orientation and connect the connector.
4. Install the HLX-801 inside the maintenance compartment.
At this time, follow the precautions below when inserting the HLX-801.
・
Be careful not to crush the connection cable.
・
Pay attention to the connector orientation. Make sure that the rating label is facing upward.
5. Attach the cover accompanying the HLX-801 instead of the maintenance cover removed in
1.
7
-
7
Содержание DSL-8001
Страница 4: ...Blank Page ...
Страница 39: ...2 1 Specification Chapter 2 Specification Performance 2 2 Specification 2 2 Performance 2 3 ...
Страница 44: ...2 6 Blank Page ...
Страница 60: ...3 16 Blank Page ...
Страница 61: ...Wiring Diagram Chapter 4 Block Wiring Diagram 4 2 4 1 ...
Страница 62: ...Block Wiring Diagram This section describes the block wiring diagram of the DSL 8001 Block Wiring Diagram 4 2 ...
Страница 64: ...4 4 Blank Page ...
Страница 72: ...6 6 Blank Page ...
Страница 102: ...7 30 Blank Page ...
Страница 148: ...11 6 Blank Page ...
Страница 150: ...3 39 4 Hongo Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 Printed in Japan 4R0100780 201301 ...