Chapter 12 Daily and Periodic Checks
12-5
12-3-1 Safety Check (6 items)
y
In order to ensure safe handling of medical equipment and measuring instruments and to enable tests to be
performed safely, leakage current and other factors that have adverse effect on human body are prescribed
by the safety test standards.
When performing the check, assessment should be made by measuring the leakage current with a leakage
current measure that satisfies the measurement conditions for medical electrical equipment.
WARNING
If, during a safety inspection, a value that exceeds the allowable limit is confirmed,
immediately stop using the given unit; then replace the defective part or make
proper repairs. Failure to repair the unit can lead to a serious accident.
Note:
z
There are a several types of measurement states in the measurement of leakage current, be
sure to determine the maximum value among all the measured values and adopt it as the
measured value. It is also important to compare the measured values against those in the
past and check whether there are any problematic changes in them.
Содержание CardiMax FX-8322
Страница 4: ......
Страница 16: ......
Страница 17: ...Chapter 1 Overview C Ch ha ap pt te er r 1 1 Overview 1 Overview 1 1 ...
Страница 18: ...Chapter 1 Overview ...
Страница 20: ...Chapter 1 Overview 1 2 ...
Страница 22: ...Chapter 2 Specifications ...
Страница 28: ...Chapter 2 Specifications 2 6 ...
Страница 30: ...Chapter 3 Names and Functions of Parts ...
Страница 34: ...Chapter 4 Operational Description ...
Страница 48: ...Chapter 5 Internal Connection Diagram ...
Страница 50: ...Chapter 5 Internal Connection Diagram 5 2 ...
Страница 51: ...Chapter 6 Spare Parts List C Ch ha ap pt te er r 6 6 Spare Parts List 6 Spare Parts List 6 1 ...
Страница 52: ...Chapter 6 Spare Parts List ...
Страница 56: ...Chapter 6 Spare Parts List 6 4 ...
Страница 58: ...Chapter 7 Disassembly and Assembly Procedure ...
Страница 75: ...Chapter 7 Disassembly and Assembly Procedure 7 17 7 17 Outside Drawing ...
Страница 76: ...Chapter 7 Disassembly and Assembly Procedure 7 18 ...
Страница 78: ...Chapter 8 Software Update Procedure ...
Страница 81: ...Chapter 9 Caution Labels C Ch ha ap pt te er r 9 9 Caution Labels 9 Caution Labels 9 1 ...
Страница 82: ...Chapter 9 Caution Labels ...
Страница 84: ...Chapter 9 Caution Labels 9 2 ...
Страница 86: ...Chapter 10 Troubleshooting ...
Страница 98: ...Chapter 11 Maintenance ...
Страница 114: ...Chapter 12 Daily and Periodic Checks ...
Страница 125: ...Chapter 12 Daily and Periodic Checks 12 11 3 Touch Maintenance 4 Touch Manufacturer maintenance 3 4 ...
Страница 140: ......
Страница 142: ...39 4 Hongo 3 chome Bunkyo ku Tokyo Japan Phone 81 3 3815 2121 Fax 81 3 3814 1222 4R0100320 201003 ...