DICOM configuration pages
In order to archive exams, patient data and images, you must first configure the
DICOM settings.
You must also configure the Vevo MD Imaging System for DICOM transfer. Have
your Network Administrator or PACS Manager perform the configuration task. See
"Configuring the system for DICOM transfer"
Important:
Changing the configuration requires a system restart.
The DICOM configuration pages are as follows:
l
Location
Configuration for network settings. Also specifies which devices you
want to associate with that network. For example, configure a location called
“Office,” and then associate an archiver with it. You can configure up to eight
locations. See
"Associating devices with locations"
l
Archive
Configuration for PACS archivers, devices for storing patient exams
(images and clips). Only one archiver per location can receive image transfers
at a time. You can associate up to four archivers per location.
l
Storage Commit
Configuration for storage commitment servers, devices that
take responsibility for and provide receipt of content sent by the ultrasound
system.
l
Worklist
Configuration for worklist servers, devices that contain scheduled
patient procedure data. You can import the data into the patient form, see
l
MPPS
or Modality Performed Procedure Step. Configuration for MPPS servers
that coordinate detailed information about the exam performed. You can
associate one MPPS server per location.
82
System settings
Содержание VisualSonics Vevo MD
Страница 1: ......
Страница 2: ......
Страница 12: ...12 ...
Страница 27: ...System enclosure with composite safety warning label location highlighted System components 27 ...
Страница 67: ...Contraindications The Vevo MD Imaging System system has no known contraindications Contraindications 67 ...
Страница 69: ...System settings 69 ...
Страница 70: ...70 System settings ...
Страница 77: ...3 Tap DICOM Setup Connectivity 77 ...
Страница 112: ...3 Tap Delete and confirm you want to delete the exam type 4 Tap Done to close the settings page 112 System settings ...
Страница 114: ...4 Type a name for the exam type Note The exam type must have a unique name 5 Tap Done 114 System settings ...
Страница 146: ...2 Tap User Management in the list on the left 146 System settings ...
Страница 164: ...2 Check the Enable Patient Header check box below the user list 3 Tap Done 164 System settings ...
Страница 168: ...Review images screen 1 Next and previous image 2 Scan 3 Export 4 Delete 168 Patient ...
Страница 174: ...3 Tap a text box and fill in the field See Patient form fields on page 170 4 Tap Done 174 Patient ...
Страница 266: ...Note When you enable zoom from inside the color box the whole image is zoomed not just the color box area 266 Scanning ...
Страница 281: ...3 When the move is complete lift your finger to commit the change 4 Save your image Annotations 281 ...
Страница 293: ...4 Tap Save Clip to save the entire clip or tap Save Frame to save a new single frame To add a linear measurement 293 ...
Страница 461: ...zoom while scanning 2D control 226 Color Doppler Mode control 265 Index 461 ...
Страница 462: ...462 Index ...
Страница 463: ...51370 01 1 0 51370 01 ...