Storing the transducer
113
6 Clean
the
TRANSDUCER CABLE AND BODY
to remove any debris or bodily fluids. Use the following
procedure:
a
Use
either a premoistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner
Table 7-2, “Cleaners and disinfectants compatible with M-Turbo system and transducers”
b
Remove all gel, debris, and bodily fluids from the transducer.
c
With a new wipe, clean the cable and transducer, starting from the cable, wiping toward the scanhead.
This method helps to avoid cross-contamination.
d
Observe the minimum wet contact time. Monitor the transducer for wet appearance. Re-apply with a
new wipe if no longer wet.
7 Verify
that all gel, debris, and bodily fluids have been removed from the system and transducer. If
necessary, repeat steps 5 and 6 with a new wipe.
8
Allow
the ultrasound system, including the console and transducer, to air dry in a clean, well-ventilated
space.
9
Examine
the system, transducer and cable for damage, such as cracks or splitting where fluid can enter.
If damage is evident, do not use the transducer. Instead, contact FUJIFILM SonoSite or your local
representative.
Storing the transducer
To store the transducer
1
Make sure the transducer has been cleaned and disinfected as detailed in this chapter.
2
Store the transducer so that it hangs freely and vertically, and observe the following precautions:
Store the transducer away from any contaminated transducers.
Caution
Do not allow moisture near the electronic components of the connector.
Содержание SonoSite Edge
Страница 1: ...Cleaning and Disinfecting FUJIFILM SonoSite Products User Guide...
Страница 24: ...18 Compatible high level disinfectants compatible transducers...
Страница 46: ...40 Compatible high level disinfectants transducers...
Страница 68: ...62 Compatible high level disinfectants transducers...
Страница 88: ...82 Compatible high level disinfectants transducers...
Страница 110: ...104 Compatible high level disinfectants transducers...
Страница 150: ...144 Compatible high level disinfectants transducers...
Страница 151: ......
Страница 152: ...P22043 01 P22043 01...