9 Function Menu
9.3.20 Test Item Name Setting
You can configure the display of item names. The names appear in the screen display and printed
paper of measurement results.
1
2
3
1
Choose the item names to be used for display
and printing.
2
Choose the names to be used for external com-
munication output to the host computer.
The selection of [ALPi, LDHi]/[ALP, LDH] is the
display setting of the test name ALPi, LDHi.
This is not the display setting of the test name
ALP, LDH.
(ALPi and LDHi are slides compatible with the
IFCC method. Availability depends on the re-
gion of sale.)
NOTE
If [ALP, LDH] is selected, ALPi, LDHi ap-
pears the same as ALP, LDH.
Check this setting again before testing.
3
Choose the item names to be used. Tap either
the [BUN] or [UREA] button.
NOTE
When the name of BUN is changed to
UREA, or vice versa, set again the fol-
lowing items.
ITEMS
Reference interval
Correlation coefficient
Control fluid settings
NOTE
When BUN/UREA is changed, the change
is reflected in the following items:
• Test Result
• Unit conversion
• Reference Interval setting
• Calculated Parameter
•
Dilution factor Coefficients
NOTE
Regarding the BUN ↔ UREA conversion
coefficient
UREA = BUN×2.14 [mg/dL]
UREA = BUN [mmol/L]
NOTE
Rebooting the device is necessary for the setting to take effect. The confirmation dialog for
the power-off will appear when
is tapped to exit Mode function for administrator. After the
shutdown, reboot the analyzer.
9
Содержание FUJI DRI-CHEM NX600
Страница 2: ......
Страница 33: ......
Страница 72: ......
Страница 106: ......
Страница 160: ......
Страница 167: ...8 Troubleshooting 8 3 897N203029A FDC NX600 Instruction Manual 2022 01 8...
Страница 248: ......
Страница 250: ......
Страница 254: ...11 USM User Safety Mode 11 4 897N203029A FDC NX600 Instruction Manual 2022 01 ADM All functions are operable 11...
Страница 266: ......
Страница 267: ......