16-5
16.1 Replenisher Cartridge Section
16
4. Disconnect the connector, release the locking
tabs and remove the sensor.
Installation
Installation is essentially in the reverse order of removal.
Removal
1. Remove:
• Move the water supply port assembly to one
side
• Replenisher box door open/close detecting
interlock switch bracket
• Replenisher cartridge setting sensor bracket
• Replenisher cartridge box upper/lower sensor
bracket
.
2. Open the four clamps at the right-hand side and
six clamps at the left-hand side.
3. Disconnect the motor connector and open the
clamp.
LII687
Connector
Locking Tab
Sensor (D722)
Sensor (D723)
16.1.4
Replenisher Cartridge Open Unit Removal/Reinstallation
LII684
Clamps (4)
Clamps (6)
Connector
Clamp