15-14
15.2 Processing solution Heater/Circulation Pump Section
1. Connect the hoses securely so that they extend
as far as the base of the connection opening.
2. Inspect the hose clamps, and replace them if the
docking portion is damaged.
3. Check that the hose clamps are not twisted when
they are installed. Hold each end of the clamp
with a pair of pliers and squeeze the ends
together securely.
4. After operating the machine for 30 minutes, check
that there is no solution leakage from the hose
connections.
15.2.9
Hose Connection
0681
Docking Portion
Hose Clamp
0680
Hose Clamp