5-115
5.6 Printer Adjustment/Maintenance (05)
5
This procedure calibrates the paper sensors.
[Selection]
[Setup and Maintenance]
→
[05 Printer Adjustment/
Maintenance]
→
[0558 Sensor Calibration (Initial)]
D619 : Upper magazine splice sensor
D629 : Lower magazine splice sensor
D650 : Exposure section entrance paper sensor.
D653 : Exposure position paper sensor
1. Click the [
▼
] button to select the sensor to be
calibrated.
2. Click the [Start] button.
• Date and result are displayed.
NOTE: If the result is “NG”, the sensor should be
replaced.
3. Click the [Close] button.
• Operation returns to the menu screen.
5.6.25
Sensor Calibration (Initial) (0558)