5-57
5.4 Scanner Adjustment/Maintenance (03)
5
• The “I-4389” message appears.
4. Click the [OK] button.
5. Click the [Close] button.
• Operation returns to the menu screen.
6. Perform the steps in the following menus.
- 0347 “Focus Calibration”
- 0321 “Focus Position Adjustment” for each
carrier and mask
The center value for each printing magnification can
be set up in this menu.
The NC100AY with film in the 135-lane can be used
to execute this menu if the focusing chart jig is not
available.
Set a film in the 135F mask for the MFC10AY.
[Selection]
[Setup and Maintenance]
→
[03 Scanner Adjustment/
Maintenance]
→
[0347 Focus Calibration]
1. Install the focusing chart jig instead of the carrier.
5.4.13
Focus Calibration (0347)
z
2
212
Focusing Chart Jig