107
FMPC
8.2.29 Processor Temperature
Calibration (467)
[Selection]
“4 Setup and Maintenance” - “6 Processor
Adjustment/Maintenance” - “7 Processor
Temperature Calibration”
[Procedure]
Perform the processor section temperature
calibrations.
1
Check that the display values reach the “Setting”
temperatures.
2
Turn the P1 filter holder 90° counterclockwise,
remove the P1 circulation filter and place it on
the tray.
NOTE:
Do not let the solution drip or run.
3
Measure the temperature of the processing
solution and record it.
4
Reinstall the filter by aligning the fastening tab
with the notch, and turn it 90° clockwise.
5
If the “Setting” temperatures differ from the
values displayed, move the cursor to the
“Measured value” and enter the values recorded
in Step 3.
Range : 0.0 to 99.9
6
Click [Enter].
7
For P2, PS2 and PS4, perform Steps 1 through
6.
P1 Filter Holder
EZ680
EZ749
Fastening Tab
Notch
EZ682