A-1
Appendix
A Specifications
DR-ID 670 Operation Manual
897N102735
Appendix A
Specifications
A.1 Specifications
Specifications of the DR-ID 670 are shown below.
A.1.1 Reduced Equivalent (DR-ID 670)
Peak reduced equivalent on the front panel of the flat panel sensor: 0.5 mmAl
A.1.2 Power Supply Conditions
Rated voltage : 100-240V ±10% ~
Input current
: 1-0.42A
Frequency
: 50-60Hz
A.1.3 Environmental Conditions
(1) Operating Conditions
Temperature
: 15°C (15%RH) - 30°C (80%RH)
Humidity
: 15%RH (15°C) - 80%RH (30°C) (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
(2) Non-operating Conditions
(Environmental conditions under which power can be supplied)
Temperature
: 5°C - 35°C
Humidity
: 10%RH - 80%RH (no dew condensation)
Atmospheric pressure : 700hPa - 1060hPa
CAUTIONS
When the flat panel sensor is used in high temperature condition for long period of time, it may
cause image artifacts and/or failure of the device.
A.1.4 Image Performance
Each flat panel sensor complies with IEC 62220-1 (MEDICAL ELECTRICAL EQUIPMENT -
CHARACTERISTICS OF DIGITAL X-RAY IMAGING DEVICES - ) as a general X-ray radiography
equipment.
To ensure optimal image quality, it is recommended that you do not use the flat panel sensor near
devices (motor, transformer, switching supply, etc.) that generate electromagnetic noise.
Содержание DR-ID 670
Страница 2: ...ii DR ID 670 Operation Manual 897N102735...
Страница 24: ...1 18 For Safe Operation 1 DR ID 670 Operation Manual 897N102735...
Страница 32: ...2 8 System Configuration Product Overview 2 DR ID 670 Operation Manual 897N102735...
Страница 42: ...4 2 Daily Inspection and Maintenance 4 DR ID 670 Operation Manual 897N102735...
Страница 44: ...A 2 Appendix A Specifications DR ID 670 Operation Manual 897N102735...
Страница 54: ...Maintenance and Inspection DR ID 670 Operation Manual 897N102735...
Страница 61: ...Maintenance and Inspection DR ID 670 Operation Manual 897N102735...
Страница 62: ......