1-12
For Safe Operation
1
DR-ID 670 Operation Manual
897N102735
Table 3
Guidance and manufacturer’s declaration - electromagnetic immunity
The DR-ID 670 is intended for use in the electromagnetic environment specified below.
The customer or the user of the DR-ID 670 should assure that they are used in such an environment.
Immunity test
IEC 60601-1-2 test level Compliance level
Electromagnetic environment - guidance
Conducted RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the DR-ID
670, including cables, than the recommended
separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance
d
= 1.2
d
= 1.2
80 MHz to 800 MHz
d
= 2.3
800 MHz to 2.5 GHz
where
P
is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and
d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,
a
should be less than the compliance level in each
frequency range.
b
Interference may occur in the vicinity of equipment
marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
a Field strength from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the DR-ID 670 is used exceeds the
applicable RF compliance, the DR-ID 670 should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the DR-ID 670.
b Over the frequency range 150 kHz to 80 MHz, field strength should be less than 3 V/m.
Содержание DR-ID 670
Страница 2: ...ii DR ID 670 Operation Manual 897N102735...
Страница 24: ...1 18 For Safe Operation 1 DR ID 670 Operation Manual 897N102735...
Страница 32: ...2 8 System Configuration Product Overview 2 DR ID 670 Operation Manual 897N102735...
Страница 42: ...4 2 Daily Inspection and Maintenance 4 DR ID 670 Operation Manual 897N102735...
Страница 44: ...A 2 Appendix A Specifications DR ID 670 Operation Manual 897N102735...
Страница 54: ...Maintenance and Inspection DR ID 670 Operation Manual 897N102735...
Страница 61: ...Maintenance and Inspection DR ID 670 Operation Manual 897N102735...
Страница 62: ......