10.5.2 Contact input setting
Description
You can set the functions for the contact inputs 1 to 3 using this function.
Settable range: Select one of the following
(1)
[NONE]
: Performs no treatment by contact input.
(2)
[BLOW DOWN ON]
:
Performs blowdown by contact input.
(Switch OFF to ON to perform blowdown.)
(3)
[HEATER OFF]
:
Turn off the heater by contact input.
(OFF
/
ON
:
Heater ON
/
Heater OFF)
(4)
[PROHIBIT CAL.]
:
Sets if calibration is prohibited or valid by contact in-
put.
(OFF/ON: Calibration is valid/prohibited.)
(5)
[REMOTE CAL.]
:
Performs all calibration by contact input.
(Switch OFF to ON to perform calibration.)
(6)
[REMOTE HOLD]
: Holds the AO by contact input.
(OFF/ON: not held/held)
(7)
[CALCULATE REST]
:
Resets maximum and minimum calculations of O
2
by
contact input.
(Switch OFF to ON to perform calibration.)
(8)
[OUTPUT RANGE]
:
Switches the range by contact input.
(OFF
/
ON: Range-1/Range-2)
Note) The functions other than “NONE” cannot be set for multiple contacts.
Procedu
re
Operation
(example)
Setting the blowdown function for the contact input 1
Key operation
Description
Displayed message
(LCD)
(1)
Display the screen on the right in accordance with the key op-
eration summary and press the
key.
PARAMETER MENU
DIGITAL INPUT
(2)
Use the
key to select the contact input 1 setting screen.
Press the
key to set the value.
(Also follow this procedure for the contact inputs 2 and 3.)
DIGITAL INPUT
DI 1
(3)
Use
the
key to select the function for contact input 1.
DI 1
NONE
(4)
The item selected is highlighted.
Press the
key to set the value.
DI 1
BLOW DOWN ON
(5)
Press
the
key.
DI 1
BLOW DOWN ON
(6)
The display returns to the screen on the right.
DIGITAL INPUT
DI 1
Содержание ZKMA
Страница 1: ......
Страница 13: ...2 INZ TN2ZKMA E 2 OPERATING PARTS AND THEIR FUNCTIONS 2 1 Outline Drawing 1 ZKMA 2 ZKMB...
Страница 22: ...11 INZ TN2ZKMA E 4 2 2 Cable gland and input and output lines 1 ZKMA 2 ZKMB...
Страница 23: ...12 INZ TN2ZKMA E 4 2 3 Allocation of the terminal block...
Страница 24: ...13 INZ TN2ZKMA E 4 3 Wiring and piping diagram 4 3 1 1 Flow guide tube system...
Страница 25: ...14 INZ TN2ZKMA E 4 3 1 2 Flow guide tube system with valve...
Страница 26: ...15 INZ TN2ZKMA E 4 3 1 3 Ejector system...
Страница 27: ...16 INZ TN2ZKMA E 4 3 1 4 Ejector system with valve...
Страница 31: ...20 INZ TN2ZKMA E 5 2 Key operation flow diagram outline...
Страница 32: ...21 INZ TN2ZKMA E...
Страница 33: ...22 INZ TN2ZKMA E...
Страница 34: ...23 INZ TN2ZKMA E...
Страница 113: ...102 INZ TN2ZKMA E 11 SPECIFICATIONS 11 1 Specifications...
Страница 114: ...103 INZ TN2ZKMA E...
Страница 115: ...104 INZ TN2ZKMA E 11 2 Code symbols Converter...
Страница 116: ...105 INZ TN2ZKMA E Detector...
Страница 117: ...106 INZ TN2ZKMA E Ejector Exclusive cable...
Страница 118: ......