Preface

We  thank  you  very  much  for  purchasing  Fuji’s  NO

2

/NO 

Converter.

●  Please read this instruction manual thoroughly before installing, 

using and applying maintenance this device.

  Damage or accident may be caused when mishandled.

●  Specification of this NO

2

/NO converter may alter without 

notice for modification.

●  Do not remodel or modify this NO

2

/NO converter without the 

manufacturer’s permission. Fuji Electric will not accept any 
liability whatsoever for any trouble or accident caused by such 
modification.

●  Operator of the NO

2

/NO converter should keep this instruction 

manual.

●  Operator should keep this instruction manual near at hand at all 

times, after thoroughly reading it.

Safety precautions

Please read the safety precautions written as bellow before 
use, for correct use of the converter.

Please observe cautions stated bellow, for it contains important 
information on safety.

  Installation,  wiring  and  piping  should  be  carried  out  by 

professionals or suppliers. Incomplete installation may cause 
fall of the device, electric shock or fire.

  Gas analyzer should be turned OFF when wiring, maintenance 

or inspection is carried out. This is to prevent electric shock and 
injury.

  Use wires with proper wire rods and diameters that meets this 

device. Wrong ones may cause electric shock or fire.

  Do not insert metal rod or fingers to the power supply terminal. 

It may cause electric shock.

  Remove any metal objects such as wristwatch while operating 

maintenance or inspection. It may cause electric shock.

  Use  pipes  with  proper  material  and  joints  stated  in  the 

instruction manual. It may cause gas leakage.

  Use replacement parts that are specified by the manufacturer. 

Otherwise, it may cause malfunction, electric shock or gas 
leakage.

  Do not touch the converter section, due to high temperature. 

Wear protective gloves while exchanging catalyst. It may cause 
burn injury.

1. Overview

The NO

2

/NO converter is to be coupled with a NOx gas 

analyzer  or  NH

3

  gas  analyzer  for  flue  exhaust.  It  is 

converter  using  a  special  catalyst  which  efficiently 
converts NO

2

 in sample gas to NO.

2. Major Specifications

Catalyst 

:  Carbon,  replacement,  required 

every 8 months

 

  (when  NO

2

  concentration  is  10 

ppm or lower)

Gas flow rate 

:  About 0.5L/min.

Set temperature 

:  220±10°C

Thermocouple 

:  K

Power supply 

:  100 to 240V AC, 50/60Hz

Power consumption  :  About 85VA

Altitude 

:  Up to 2,000 m

Installation category :  II

Pollution Degree 

:  2

Gas inlet/outlet connection method: 

 Insert Teflon  tube  ø6  mm/ø4  mm  into Viton 

connection  port  of  inner  diameter  ø5.5  mm. 

(Withstand pressure: 10 kPa)

Contact output 

:  for temperature alarm 

SPST-NC contact, 220 V AC / 30 

V DC, 1A (resistive load)

3. Outline Drawing

This gas converter is composed of the converter and the 
temperature controller.

Maintenance space

Approx. 120

(Upper side)

Maintenance space

RIGHT SIDE VIEW

Approx. 120 (Lower side)

FRONT VIEW

(NP)

OUTLET

INTLET

(NP)

124

135

148

Outlet

ø9.5/5.5

(VITON)

130

4-ø4.5

160

105

27

(212)

100 to 240V AC

Power supply

terminal

220V AC

30V DC, 1A  

Contact output (1b)

Earth terminal

Inlet

ø9.5/5.5

(VITON)

(26)

Fig. 1  Outline drawing

INSTRUCTION MANUAL

NO

2

 / NO Converter

Model: ZDL05