Preface 

We thank you very much for purchasing Fuji’s NO

2

/NO 

Converter. 

 

  Please read this instruction manual thoroughly 

before installing, using and applying maintenance 
this device.   
Damage or accident may be caused when 
mishandled. 

  Specification of this NO

2

/NO converter may alter 

without notice for modification.   

  Do not remodel or modify this NO

2

/NO converter 

without the manufacturer’s permission. Fuji Electric 
will not accept any liability whatsoever for any 
trouble or accident caused by such modification.   

  Operator of the NO

2

/NO converter should keep this 

instruction manual.   

  Operator should keep this instruction manual near at 

hand at all times, after thoroughly reading it. 

 

Safety precautions

 

Please read the safety precautions written as bellow 
before use, for correct use of the converter. 

Please observe cautions stated bellow, for it contains 
important information on safety. 
 
 
 
 
 
 

CAUTION

 

 
 
 
 
 
 

 
 
 
 
 
 
 
 
 
 

CAUTION

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

1. Overview

 

The NO

2

/NO converter is to be coupled with a NOx gas 

analyzer or NH

3

 gas analyzer for flue exhaust.    It is 

converter using a special catalyst which efficiently 
converts NO

2

 in sample gas to NO. 

 

2. Major Specifications

 

Catalyst 

:  Carbon, replacement required every 

8 months

 

(when NO

2

 concentration 

is 10 ppm or lower) 

Gas flow rate 

:  About 0.5L/min. 

Set temperature 

:  220±10°C 

Thermocouple : 

Power supply 

:  100V AC, 50/60Hz 

Power consumption  :  About 85VA 
Material of parts that contact gas 
 

:  Ceramic, Viton, Glass wool, 

SUS316 

Gas inlet/outlet connection method 
 

:  Insert Teflon tube 

φ

6 mm/

φ

4 mm 

into Viton connection port of inner 
diameter 

φ

5.5 mm. (Withstand 

pressure: 10 kPa) 

NO

/ NO Converter 

INZ-TN1ZDL03-E 

INSTRUCTION MANUAL

  Installation, wiring and piping should be carried 

out by professionals or suppliers. Incomplete 
installation may cause fall of the device, electric 
shock or fire.   

  Gas analyzer should be turned OFF when wiring, 

maintenance or inspection is carried out. This is to 
prevent electric shock and injury.

 

  Use wires with proper wire rods and diameters 

that meets this device. Wrong ones may cause 
electric shock or fire.   

  Do not insert metal rod or fingers to the power 

supply terminal. It may cause electric shock.   

  Remove any metal objects such as wristwatch 

while operating maintenance or inspection. It may 
cause electric shock. 

  Use pipes with proper material and joints stated in 

the instruction manual. It may cause gas leakage.   

  Use replacement parts that are specified by the 

manufacturer. Otherwise, it may cause 
malfunction, electric shock or gas leakage.   

  Do not touch the converter section, due to high 

temperature. Wear protective gloves while 
exchanging catalyst. It may cause burn injury.

 

Model: ZDL03

Precautions under this mark is stated 
when wrong handling may cause 
hazardous situation. Possibility of 
Medium level damage or injury, and 
physical damage is predicted. 

CAUTION