DigiTemp User Manual 

Chapter 2  SDI Commands 

 

page 10 

2.3.1

 

Address Query 

This command requests the address of the SDI sensor. 

 

String 

Note 

Cmd 

?! 

request the (single) device on this bus to report its address 

Resp 

the device is configured for address 0 
only one SDI device should be connected to the bus when 

using this command; problems arise when several devices 

respond on the same bus 

2.3.2

 

Acknowledge Active 

This command queries whether a sensor is present on the SDI bus at the specified address. 

 

String 

Note 

Cmd 

0! 

request the device at address 0 to confirm it is active 

Resp 

a device is present at address 0 

2.3.3

 

Change Address 

This command changes a sensor’s SDI address. 

 

String 

Note 

Cmd 

0A3! 

change the address of the device at SDI address 0 to 3 

Resp 

response confirms change 

2.3.4

 

Send Identification 

This command requests detailed identification information from the addressed sensor. 

 

String 

Note 

Cmd 

3I! 

 

Resp 

313FTS-----DgiTmp-v134567  3

 

device SDI address 

13

 

compatible with SDI-12 version 1.3 

FTS

 

 manufacturer’s identifier 

DgiTmp

 

sensor model 

-v1

 

version 1 of sensor firmware 

34567

  sensor serial number 

2.4

 

SDI Data commands 

SDI data commands request data from the sensor. 

Содержание DigiTemp

Страница 1: ...1 800 548 4264 www ftshydrology com 700 DigiTemp Rev 1 0 24 Nov 2011 DigiTemp SDI 12 Submersible Digital Temperature Sensor User Manual...

Страница 2: ...Contact information FTS 1065 Henry Eng Place Victoria B C V9B 6B2 CANADA www ftshydrology com Toll free 1 800 548 4264 Local 250 478 5561 Technical support support ftshydrology com...

Страница 3: ...Revision History Date Person Description 24 Nov 2011 RG First released version 28 Nov 2011 RG Late corrections...

Страница 4: ...3 Connection 7 1 4 Operation 7 1 5 Configuration 7 1 6 Calibration and maintenance 7 Chapter 2 SDI Commands 8 2 1 Notation for SDI commands 9 2 2 Commands recognized by DigiTemp 9 2 3 General SDI com...

Страница 5: ...Chapter 1 Operation Chapter contents 1 1 General description 6 1 2 Installation 6 1 3 Connection 7 1 4 Operation 7 1 5 Configuration 7 1 6 Calibration and maintenance 7...

Страница 6: ...these steps 1 Place the sensor in a location where it is continuously immersed in a representative sample of the water to be monitored usually a well mixed locale 2 The sensor may be mounted on an ob...

Страница 7: ...s powered by the 12 V DC power supply on the SDI 12 bus For details on the SDI commands used by the sensor see Chapter 2 SDI Commands 1 5 Configuration The DigiTemp is shipped with default address 0 u...

Страница 8: ...2 3 General SDI commands 9 2 3 1 Address Query 2 3 2 Acknowledge Active 2 3 3 Change Address 2 3 4 Send Identification 2 4 SDI Data commands 10 2 4 1 Start Measurement 2 4 2 Start Concurrent Measureme...

Страница 9: ...Commands recognized by DigiTemp The table below provides a quick reference to the commands recognized by the DigiTemp For details on these commands see the following sections Command name Command code...

Страница 10: ...address String Note Cmd 0 request the device at address 0 to confirm it is active Resp 0 a device is present at address 0 2 3 3 Change Address This command changes a sensor s SDI address String Note C...

Страница 11: ...Table 2 String Note Cmd 3C or 3CC start a concurrent measurement on sensor at address 3 without or with CRC in data response respectively Resp 300302 3 device SDI address 003 measurement delay until d...

Страница 12: ...ata is returned Cmd 1RC perform a continuous measurement read on sensor at address 3 with CRC code in data response Resp 1MSA 1 sensor SDI address MSA CRC code DigiTemp is not a continuous measurement...

Страница 13: ...Chapter 3 Technical specifications...

Страница 14: ...may be less accurate Responsiveness Time constant 0 57 min reaches 95 of final value in 1 7 min 99 in 2 9min Measurement time 3 s Power supply voltage 12 V DC nominal range 9 6 V 18 6 V Standby curre...

Отзывы: