Section 6
Parameter Settings
68
FT205EV Wind Sensors
– User Manual
6.4.24 UC.2: Clear User Calibration Table Record
Command
Parameter
UC (Erase table)
Command
Syntax
SET Sensor:
$<listenerID>,UC<erase>*<checksum><cr><lf>
$aa,UCCLEAR*hh<cr><lf>
Parameters
<erase>
CLEAR
Erases both Flash and RAM copies of user calibration table
Examples
Example 1
Erase the user calibration tables and verify
Message
Comment
$01,UCCLEAR*62<cr><lf>
Erase calibration tables
$01,UC?*04<cr><lf>
Query user calibration table status
$WI,UC=00,D,0000,0000*71<cr><lf>
Sensor response
Description
Use the UCCLEAR command to erase the RAM and saved FLASH copies of the user
calibration table. A UCCLEAR is performed before a new user calibration table is loaded (see
Section 6.4.25).
The user calibration table label is also cleared to 32 ASCII spaces when the UCCLEAR
command is sent (see Section 6.4.27).