6.5 Clamping the grinding bowls
6.5.1 Clamping with the "Safe-Lock" (5) tensioning device
Fritsch GmbH confirms that every Safe-Lock clamping
system has been manufactured and tested to our
internal quality standards. Fritsch GmbH measures the
clamping force of each individual Safe-Lock clamping
system. The clamping force must be in the range of 11.2
- 11.3 kN +/- 0.2kN.
Click on the following link to watch a step-by-step video guide on how to
use the Safe-Lock clamping device. Click here.
Prior to every grinding, carry out the following checks before clamping
the grinding bowls:
n
Has the rubber disk (6.1) been inserted into the grinding bowl
holder? The rough side should be facing upwards!
n
Check the rubber disk in the grinding bowl holder for damage:
Replace the rubber disk if it has been pressed flat!
n
The Teflon flat seal (for the sealing between lid and bowl) may not be
damaged or soiled.
Replace heavily deformed Teflon flat seals.
n
The surfaces of the lids and bowls on which the Teflon flat seals lie
must be clean!
n
Check the rubber disk of the pressure piece for damage!
Replace rubber disks that have been pressed flat and are protruding
laterally from the pressure piece.
n
Make sure that the thread on the Safe-Lock is not soiled and not
damaged. Both can lead to a premature actuation of the torque
spanner and thus preventing the required final clamping force from
being reached.
Clamping
1.
Position the bracket (a) of the clamping lever vertically!
Using the device