Checklist for troubleshooting
Fault description
Cause
Remedy
No grinding noise or device
switches off completely
although there is material in
the grinding chamber.
Sample blocking
crushing plates
n
Enlarge gap width to loosen sample.
n
Remove guide pin of fixed crushing jaw and remove fixed
crushing jaw.
n
If neither option is possible, you can loosen the sample
from below using a screwdriver, for example.
Final fineness becoming
larger
Gap width becoming
larger due to abra-
sion or use
Check gap width after several grinding processes and read-
just, if necessary. See
Chapter 6.3.3 ‘Checking the gap
Checklist for troubleshooting
- 48 -