![Fritsch ANALYSETTE 3 Скачать руководство пользователя страница 27](http://html.mh-extra.com/html/fritsch/analysette-3/analysette-3_operating-instructions-manual_2338411027.webp)
6.1.3 Feed quantity for dry or wet sieving
Maximum load:
‐ Sieving stock < 1 kg
‐ Sieves and sieving stock < 6 kg
6.2 Dry sieving
For dry sieving, use the dry sieve pan (without outlet) and the dry
clamping lid (plexiglas clamping lid without nozzles).
6.2.1 Sieving parameters
Parameters
Coarse sample
Fine sample
Sieving time
3...20 min
15....30 min
Amplitude
2.5....3 mm
1.5....2.5 mm
In order to achieve an amplitude set point of 3 mm (full deflection), at
least 3 sieves, collecting vessels and clamping lids must be clamped on
the sieve shaker. If fewer sieves are clamped, it may not be possible to
achieve an amplitude set point of 3 mm.
(Sieves with 200 mm diameter)
Using the device
- 25 -
Содержание ANALYSETTE 3
Страница 4: ......
Страница 53: ...14 Safety logbook Date Maintenance Repair Name Signature Safety logbook 51 ...
Страница 54: ...Date Maintenance Repair Name Signature Safety logbook 52 ...
Страница 57: ......