DANGER!
Before using other measuring liquids, consult Fritsch
Gmbh in advance or call up the list of suitable chemi-
cals on our homepage www.fritsch-sizing.de under the
corresponding wet dispersion unit.
We advise against the use of harmful, explosive or flam-
mable liquids.
The above list just states the chemical compatibility of
the device with the liquids that can be used.
The measuring unit and dispersion units are not
designed to be explosion-protected.
DANGER!
The following must not be used:
Ketones (e.g. acetone, propanone, butanone, cyclohex-
anone),
Ether, chlorofluorocarbons,
Amines, freone 21-32, methanol, aniline, benzol
Chlorinated hydrocarbons like acetic acid and its deriva-
tives, undiluted acids and bases.
DANGER!
When using harmful liquids be sure to observe the valid
safety regulations (MAC values). Where necessary, the
measuring unit and dispersion units must be set up in a
ventilated safety area.
5.3.3 Filling the measuring circuit
In order to fill the measuring circuit of the wet dispersion unit, proceed
as follows:
1.
Check whether the hose attached to the "Liquid in" connection via
the pressure regulator is connected to your water connection or a
liquid storage tank.
2.
Also check whether the outlet hose is installed.
3.
Open the tap of the water connection or the storage container.
4.
Start ISS.
Using the device
- 38 -